Influenza A/B, RSV & SARS CoV2 (COVID 19) PCR Panel
General Information
Lab Name
Influenza A/B, RSV & SARS CoV2 (COVID 19) PCR Panel
Lab Code
FLRSNC
Epic Name
Influenza A/B, RSV & SARS CoV2 (COVID-19) PCR Panel
Description
The UW Clinical Virology Laboratory in the Department of Laboratory Medicine and Pathology incorporates the modified FDA approved Panther Fusion® SARS-CoV-2/Flu A/B/RSV assay for the rapid detection and differentiation of human influenza A/B, respiratory syncytial virus (RSV) and SARS-CoV-2 (COVID-19) using real-time PCR.
The assay targets the influenza A viruses (matrix gene), the influenza B viruses (matrix gene), RSV (matrix gene) and SARS-CoV-2 (ORF1ab gene). Results are reported as Detected or NOT detected and a semi-qualitative result is provided by reporting the PCR Cycle threshold (Ct) value for positive specimens
Synonyms
2019-nCoV, COVID, COVID-19, COVID19, FLU, Influenza A, Influenza B, nCoV, PUI nCoV, Respiratory syncytial virus (RSV), SARS, SARS-CoV-2, Wuhan Coronavirus
Components
| Code | Name |
|---|---|
| FRNVSD | Influenza A/B, RSV & SARS CoV2 (COVID 19) Specimen |
| FLAPCR | Influenza A PCR Result |
| CTFLA | Influenza A Ct Value |
| FLBPCR | Influenza B PCR Result |
| CTFLB | Influenza B Ct Value |
| RSVPCR | RSV PCR Result |
| CTRSV | RSV Ct Value |
| NCVRT | SARS CoV 2 (COVID 19) Qual PCR Result |
| NCVCT | COVID 19 Coronavirus Qual PCR Ct Value |
| FRNVI | Influenza A/B, RSV & SARS CoV2 (COVID 19) Interpretation |
Interpretation
Method
Real-time Polymerase Chain Reaction
Reference Range
See individual componentsGuidelines
Ordering & Collection
Specimen Type
Collection
Specimen requirements:
- Collect swab specimens according to standard technique using a polyester-, rayon-, or nylon-tipped swab. Immediately place the swab specimen into VTM/UTM. Do not use calcium alginate swabs. Refrigerate (2-8°C) within 4 hours of collection.
- Collect Broncho alveolar lavage (BAL) and tracheal aspirates in a sterile container.
- Minimum sample requirement: 0.75 mL
Handling Instructions
Transport specimens on ice (ice Pak), store at 2 to 8°C for up to 72 hours post collection. If there is a greater than 72 hours delay before testing of the specimen, freeze at -70°C or colder and ship on dry ice. Do not freeze/thaw.
Quantity
requested: Swab in UTM
minimum: 750 µl
Processing
Send sample to DIDD- Virology Laboratory (Central Laboratory, Renton) within 4 hours of collection.
Storage stability:
- Within 4 hours of collection: Ship the sample to the testing laboratory.
- If shipping is delayed: Store at 2–8°C (with ice packs) for up to 72 hours after collection.
- If testing is delayed beyond 72 hours: Freeze at −70°C or below and ship on dry ice.
- If a freezer or dry ice is unavailable: Ship immediately at 2–8°C (with ice packs).
- Sample rejection: Samples stored at 2–8°C for more than 72 hours after collection may be rejected.
- Do not freeze/thaw samples.
Performance
LIS Dept Code
Central Lab Virology (RVIR)
Performing Location(s)
| CL |
Division of Infectious Disease Diagnostics (DIDD) - Virology
1601 Lind Ave SW |
Phone: 206-485-9380 |
|---|
Frequency
Monday through Sunday
Available STAT?
No