Herpes Simplex Seroconversion Panel by Western Blot (Paired)
General Information
Lab Name
HSV Seroconversion by WB(Pair)
Lab Code
HSWBP
Epic Name
HSV Seroconversion by WB, Paired Sera
Description
The detection of HSV1 and HSV2 IgG class antibodies by Western blot in acute and convalescent serum. For accurate seroconversion determination, the acute and convalescent samples should be drawn at least 12 -16 weeks apart.
HSV1 and HSV2 proteins from detergent lysates ("Bernstein's lysate") of HSV infected cells are separated by electrophoresis and transferred to nitrocellulose paper. The strips of paper or "blots" containing separated fixed proteins from either HSV-1 or HSV-2 are incubated with the patient's serum. Antibodies, which bind to the viral proteins, are detected by an enzyme-mediated color change. The pattern of staining on the two blots (HSV-1 and HSV-2) is dictated by the number and identity of the HSV proteins to which the patient's immune system has antibody.
This test is reported as:
- HSV1 - Positive, Negative or Indeterminate for both the acute and convalescent sera
- HSV2 - Positive, Negative or Indeterminate for both the acute and convalescent sera
Single sera or CSF, see Herpes Simplex Western Blot [HSWB]
Components
| Code | Name |
|---|---|
| ADES20 | HSV WB (Pair) Specimen (1st) |
| ACDD20 | HSV WB (Pair) Date Drawn (1st) |
| HSWAC1 | HSV1 WB Acute Result |
| HSWAC2 | HSV2 WB Acute Result |
| CDES20 | HSV WB (Pair) Specimen (2nd) |
| CVDD20 | HSV WB (Pair) Date Drawn (2nd) |
| HSWCV1 | HSV1 WB Convalescent Result |
| HSWCV2 | HSV2 WB Convalescent Result |
| IN20 | HSV WB Interp |
Interpretation
Method
Western blot
Reference Range
See individual componentsOrdering & Collection
Specimen Type
Collection
Blood: 5 mL blood in GOLD TOP serum separator tube (SST).
Also acceptable: 5 mL blood in ORANGE TOP rapid serum tube (RST) or RED TOP tube
Quantity
requested: 1 mL serum each
minimum: 1 mL serum each
Processing
Blood: Harvest sera from whole blood within 24 hours of collection. Sera may be stored at 2-8°C (ice pack) for up to 7 days. If testing will be delayed more than 7 days from collection freeze at -10°C or colder. Do not freeze/thaw.
Outside Laboratories:
Freeze samples at -10°C or colder. Ship on dry ice, do not freeze/thaw. If freezer and/or dry ice is not available, ship immediately at 2-8°C (ice pack), shipment time must not exceed 7 days from collection. Samples stored at 2-8°C greater than 7 days old will be rejected.
Unacceptable samples: Cadaver specimens or body fluids other than human sera.
Performance
LIS Dept Code
Central Lab Virology (RVIR)
Performing Location(s)
Frequency
Monday and Thursday.
Available STAT?
No