SARS-CoV-2 Antibody, IgG Immune Status (Nucleocapsid & Quant. anti-Spike)

General Information

Lab Name

SARS CoV 2 Immune Status

Lab Code

NCVIGB

Epic Name

SARS-CoV-2 Immune Status

Description

A battery of tests to determine SARS-CoV-2 immune status:

  1. The Abbott Architect SARS-CoV-2 IgG assay, run under an emergency use authorization from the FDA, is a qualitative test designed to detect IgG antibodies to the nucleocapsid protein of SARS-CoV-2.
  2. The Abbott Architect SARS-CoV-2 IgG II assay, run under an emergency use authorization from the FDA, is a quantitative test designed to detect IgG antibodies to the spike protein of SARS-CoV-2 in serum and plasma. Results are reported as AU/mL. Per manufacture’s package insert protective level is ≥50.0 AU/mL.

For ordering individual battery components:

NCVIGG see SARS CoV 2 Nucleocapsid Antibody, IgG [NCVIGG]

NVCIGQ see SARS CoV 2 Spike Antibody, IgG [NCVIGQ]

For all questions, contact Client Support Services (available 24/7):

Phone: (206) 520-4600 or (800) 713-5198
Fax: (206) 520-4903
Email: commserv@uw.edu

The test order requisition is available online

Synonyms

SARS-CoV-2 (COVID-19) Antibody, SARS-CoV-2 (COVID-19) Serology

Components

Code Name
NCVIGR SARS CoV 2 Nucleocapsid IgG Result
NCIGAU SARS CoV 2 Spike IgG Quant
NCIGGR SARS CoV 2 Spike IgG Result
NCVISI SARS CoV 2 Immune Status Interpretation

Interpretation

Method

Chemiluminesence

Reference Range

See individual components

Ref. Range Notes

Per manufacture’s package insert, the Abbott Architect SARS-CoV-2 IgG II assay designed to detect IgG antibodies to the spike protein of SARS-CoV-2 the protective level is ≥50.0 AU/mL.

Guidelines

Ordering & Collection

Specimen Type

Collection

Preferred: 5 mL blood in GOLD SST tube.
Also Acceptable: Orange RST, pearl PPT, serum from red top, plasma from EDTA tube.

Handling Instructions

Spin within 24 hours and prior to shipment.
sample is stable for 7 days at 2-8°C once separated from a clot or red blood cells, or in a gel separator tube.
If testing will be delayed more than 7 days store at -20°C or colder.

UW Medicine
Department of Laboratory Medicine
Virology- Covid Testing Lab
1602 Raymond Ave SW
Renton, WA 98057-3356
Tel: (206) 685-6656 opt 4

Quantity

requested: Entire sample
minimum: 0.5 mL serum

Processing

Centrifuge GOLD SST tube and route to Virology Central Lab (Renton)..

Centrifuge RED TOP or EDTA tube and aliquot serum/plasma into plastic aliquot tube. Route to Virology Central Lab (Renton).

Stability: Sample stable off the clot, red blood cells, or separator gel for 7 days at 2-8°C. If testing will be delayed more than 7 days store at -20°C or colder.

Performance

LIS Dept Code

Central Lab Virology (RVIR)

Performing Location(s)

CL Virology, Renton
206-685-6656

1602 Raymond Ave SW

Renton, WA 98057-3356

206-685-6656 opt 4

Frequency

Mondays and Fridays, with results available on the same day.

Available STAT?

No

Billing & Coding

CPT codes

86413, 86769

LOINC

94504-8

Interfaced Order Code

UOW4193